The Oxygen Levels Should Continuously Decrease Over Time

Continuing Education Activity

Oxygen saturation is an essential element of patient care. Oxygen is tightly regulated within the body because hypoxemia can lead to many acute adverse effects on individual organ systems. These include the brain, heart, and kidneys. Oxygen saturation is a measure of how much hemoglobin is currently bound to oxygen compared to how much hemoglobin remains unbound. At the molecular level, hemoglobin consists of four globular protein subunits. Each subunit is associated with a heme group. Each molecule of hemoglobin subsequently has four heme-binding sites readily available to bind oxygen. Therefore, during the transport of oxygen in the blood, hemoglobin is capable of carrying up to four oxygen molecules. Due to the critical nature of tissue oxygen consumption in the body, it is essential to be able to monitor current oxygen saturation. A pulse oximeter can measure oxygen saturation. It is a noninvasive device placed over a person's finger. It measures light wavelengths to determine the ratio of the current levels of oxygenated hemoglobin to deoxygenated hemoglobin. The use of pulse oximetry has become a standard of care in medicine. It is often regarded as a fifth vital sign. As such, medical practitioners must be familiar with the functions and limitations of pulse oximetry. They should also have a basic knowledge of oxygen saturation.

Objectives:

  • Describe the physiology of the oxygen saturation curve and its shifts to the left and right.

  • Describe the indications for measuring oxygen saturation.

  • Outline the clinical significance of measuring oxygen saturation.

  • Explain the importance of improving care coordination amongst the interprofessional team to enhance the delivery of care for patients with hypoxemia.

Access free multiple choice questions on this topic.

Introduction

Oxygen saturation is an essential element in the management and understanding of patient care. Oxygen is tightly regulated within the body because hypoxemia can lead to many acute adverse effects on individual organ systems. These include the brain, heart, and kidneys. Oxygen saturation measures how much hemoglobin is currently bound to oxygen compared to how much hemoglobin remains unbound. At the molecular level, hemoglobin consists of four globular protein subunits. Each subunit is associated with a heme group. Each molecule of hemoglobin subsequently has four heme-binding sites readily available to bind oxygen. Therefore, during the transport of oxygen in the blood, hemoglobin is capable of carrying up to four oxygen molecules. Due to the critical nature of tissue oxygen consumption in the body, it is essential to be able to monitor current oxygen saturation. A pulse oximeter can measure oxygen saturation. It is a noninvasive device placed over a person's finger. It measures light wavelengths to determine the ratio of the current levels of oxygenated hemoglobin to deoxygenated hemoglobin. The use of pulse oximetry has become a standard of care in medicine. It is often regarded as a fifth vital sign. As such, medical practitioners must understand the functions and limitations of pulse oximetry. They should also have a basic knowledge of oxygen saturation.

Anatomy and Physiology

One definition of oxygen consumption within the body is the product of the arterial-venous oxygen saturation differences and blood flow. The body consumes oxygen partially through aerobic metabolism. In this process, oxygen is used to convert glucose to pyruvate, liberating two molecules of adenosine triphosphate (ATP). An important aspect of this process is the oxygen-hemoglobin dissociation curve. In the blood, hemoglobin binds free oxygen rapidly to form oxyhemoglobin leaving only a small percentage of free oxygen dissolved in the plasma. The oxygen-hemoglobin dissociation curve is a plot of percent saturation of hemoglobin as a function of the partial pressure of oxygen (PO2). At a PO2 of 100 mmHg, hemoglobin will be 100% saturated with oxygen, meaning all four heme groups are bound. Each gram of hemoglobin is capable of carrying 1.34 mL of oxygen. The solubility coefficient of oxygen in plasma is 0.003. This coefficient represents the volume of oxygen in mL that will dissolve in 100mL of plasma for each 1 mmHg increment in the PO2. A formula then calculates the oxygen content, so that Oxygen Content = (0.003 × PO2) + (1.34 × Hemoglobin × Oxygen Saturation). This formula demonstrates that dissolved oxygen is a sufficiently small fraction of total oxygen in the blood; therefore, the oxygen content of blood can be considered equal to the oxyhemoglobin levels.[1]

As PO2 decreases, the percentage of saturated hemoglobin also decreases. The oxygen-hemoglobin dissociation curve has a sigmoidal shape due to the binding nature of hemoglobin. With each oxygen molecule bound, hemoglobin undergoes a conformational change to allow subsequent oxygens to bind. Each oxygen that binds to hemoglobin increases its affinity to bind more oxygen, meaning the affinity for the fourth oxygen molecule is the highest.

In the lungs, alveolar gas has a PO2 of 100 mmHg. However, due to the high affinity for the fourth oxygen molecule, oxygen saturation will remain high even at a PO2 of 60 mmHg. As the PO2 decreases, hemoglobin saturation will eventually fall rapidly, at a PO2 of 40 mmHg hemoglobin is 75% saturated. Meanwhile, at a PO2 of 25 mmHg, hemoglobin is 50% saturated. This level is referred to as P50, where 50% of heme groups of each hemoglobin have a molecule of oxygen bound. The nature of oxygen saturation becomes increasingly important in light of the effects of right and left shifts. A variety of factors can cause these shifts.

A right shift of the oxygen saturation curve indicates a decreased oxygen affinity of hemoglobin, which will allow more oxygen to be available to tissues.[2] The mnemonic, "CADET, face Right!" can help to remember factors that can lead to a right shift. Here, "CADET" stands for PCO2, acid, 2,3-diphosphoglycerate, exercise, and temperature. The hemoglobin dissociation curve shifts right with an increase in each of these factors.

A left shift of the oxygen saturation curve indicates an increase in oxygen affinity of hemoglobin, which reduces oxygen availability to the tissues. Factors that cause a left shift in the oxygen-hemoglobin dissociation curve include decreases in temperature, PCO2, acidity, and 2,3-bisphosphoglyceric acid, formerly named 2,3-diphosphoglycerate.

Indications

Due to the noninvasive nature and relative importance of pulse oximetry readings, there are very few situations that do not indicate its use. Pulse oximetry can provide a rapid tool to assess oxygenation accurately. It is particularly useful in emergencies for this reason. Cyanosis may not develop until oxygen saturation reaches about 67%. As such, pulse oximetry is extremely useful because the signs and symptoms of hypoxemia may not be visible on physical examination.

Indications for pulse oximetry include any clinical setting where hypoxemia may occur. These settings include patient monitoring in emergency departments, operating rooms, emergency medical services systems, postoperative recovery areas, endoscopy suites, sleep and exercise laboratories, oral surgery suites, cardiac catheterization suites, facilities that perform conscious sedation, labor and delivery wards, interfacility patient transfer units, altitude facilities, aerospace medicine facilities, and even patients' homes.[3]

Contraindications

There are rarely contraindications to pulse oximetry is rarely contraindicated, but understanding its limitations is helpful. A relative contraindication may be a need to measure pH, PaCO2, total hemoglobin, and abnormal hemoglobins as in the setting of carbon monoxide toxicity. It is also essential to monitor the location of the probe for changes in skin conditions such as blisters or damage to the nail bed. Patients with burns may also require the probe to be repositioned every two to four hours.

Equipment

The pulse oximeter consists of a probe containing LEDs and a photodetector. The LEDs emit light at fixed, selected wavelengths. The photodetector measures the quantity of light transmitted through a selected vascular bed such as a fingertip or earlobe. Pulse oximetry uses the Beer-Lambert law of light absorption. This law describes how light is absorbed when it passes through a clear solvent, such as plasma, that contains a solute that absorbs light at a specific wavelength, such as hemoglobin.[4] The absorption spectra of oxygenated and reduced hemoglobin differ. For this reason, arterial blood appears red, while venous blood appears blue. However, because living tissue absorbs light, it is difficult to determine the ratio of saturation of hemoglobin in the body. The oximeter probe overcomes this difficulty by emitting pulses of light, one red and one infrared. A detector is placed opposite the lights on the other side of the tissue. The diodes switch on and off in rapid sequence, and the detector measures the differences. The measurements feed into an algorithm in a microprocessor where the oxyhemoglobin saturation is calculated and eventually displayed to the user.[3]

Personnel

All medical personnel should train with a basic understanding of the use of pulse oximetry. Advanced users will find it helpful to understand the relationship of the pulse oximetry readings to blood hemoglobin concentrations, and how they are affected by the oxygen-hemoglobin dissociation curve.

Preparation

The most important consideration when preparing to apply the pulse oximeter is placing the monitor where the light will be able to shine through to the detector. Consider multiple factors before placing the pulse oximeter. Patients should remove nail polish, and the finger wiped with an alcohol preparation. Examine the finger for other objects such as excess pigmentation. For example, tattoos may block light as it passes through the tissue. High-intensity ambient light has also shown to interfere with the accuracy of pulse oximetry readings. Before application, the rubber shield should be intact on the pulse oximeter to help reduce ambient light input.

Technique

After verifying the appropriate placement site, place the pulse oximeter so that the light will penetrate through the tissue and be picked up by the detector. When placing the pulse oximeter on a fingertip, the probe must fit the finger well. It should not be too tight or too loose. Take extra caution to make sure the probe does not restrict circulation to the digit, as this may provide an inaccurate reading. There are also probes made for the earlobe. In an emergency, the pulse oximeter may have to be placed on the fingertip sideways as nail polish or pigment may obstruct the light.

Complications

Complications from using a pulse oximeter are rare. However, it is necessary to be aware of the probe site as blisters or nail damage may occur with extended use. Tissue injury may also occur in the setting of incompatible probes or during a substitution in the form of electrical shock or burns. It is also essential to know how to improve the measurements of pulse oximeters.

Possible ways to improve pulse oximeter signals include:

  • Warm and rub the skin

  • Apply a topical vasodilator

  • Try a different probe site, especially the ear

  • Try a different probe

  • Use a different machine[3]

Factors that may reduce the accuracy of pulse oximeter signals include:

  • Nail Polish[5]

  • Pigmentation of the skin

  • High-intensity ambient lighting

  • Excessive patient movement, or motion artifacts

  • Decreased perfusion

  • Presence of abnormal hemoglobin, carboxyhemoglobin

  • Intravascular dyes

  • Reduced accuracy with saturations below 83%

One significant risk of using a pulse oximeter is the possibility of treating an incorrect reading as accurate. False-negative results for hypoxemia and false-positive results for normoxemia or hypoxemia can occur. In these situations, a patient may receive inappropriate treatment, leading to harm.

False normal or high readings can occur in multiple different settings. Carboxyhemoglobin absorbs light at 660 nanometers, which is roughly the same as oxyhemoglobin. Thus, in situations where carboxyhemoglobin is high, a false normal reading may occur.[6] When glycohemoglobin A1c levels are greater than 7%, such as in patients with type 2 diabetes, an overestimation of arterial oxygen saturation may occur.[7] These situations may require an arterial blood gas to determine oxygen saturation accurately. It is also necessary to consider the clinical diagnosis when evaluating a patient with hypoxemic symptoms, as in the case of carbon monoxide toxicity.

False low readings can also occur in multiple settings. Below are some situations that may cause falsely low readings to occur.

  • Methemoglobinemia

  • Sulfhemoglobinemia

  • Sickle hemoglobin

  • Abnormal inherited forms of hemoglobin[8]

  • Severe anemia

  • Venous congestion

Clinical Significance

The human eye's ability to detect hypoxemia is poor. The presence of central cyanosis, blue coloration of the tongue and mucous membranes, is the most reliable predictor; it occurs at an oxyhemoglobin saturation of about 75%.[3] Pulse oximetry provides a convenient, noninvasive method to measure blood oxygen saturation continuously. It can also help to eliminate medical errors. Pulse oximetry has a sensitivity of 92% and a specificity of 90% when detecting hypoxia at a threshold of 92% oxygen saturation.[9]

There is no set standard of oxygen saturation where hypoxemia occurs. The generally accepted standard is that a normal resting oxygen saturation of less than 95% is considered abnormal.[10] Therefore, it remains vital to observe patients for the clinical markers of hypoxemia. The brain is the most sensitive organ, and visual, cognitive, and electroencephalographic changes develop when the oxyhemoglobin saturation is less than 80% to 85%. It is unclear whether there are long-term deficits from hypoxemia. Patients with nocturnal hypoxemia do not seem to develop life-threatening complications despite abnormally low oxygen saturation.[3]

Enhancing Healthcare Team Outcomes

All healthcare workers, including nurses, should be familiar with pulse oximetry.  Pulse oximetry is an accurate measurement of the patient's overall oxygen saturation. While few studies have demonstrated a decrease in mortality from the use of pulse oximetry, it is apparent that it provides more benefit than harm. Clinicians should be aware of the limitations and errors associated with pulse oximetry. They should use their best clinical judgment when deciding whether further workup is needed. In the case of hypoxemia, a physician should always consider whether an arterial blood sample would provide a more accurate measure of oxygen saturation than pulse oximetry.

Review Questions

This image show a Pulse Oximeter with 98% oxygen saturation level and pulse/HR of 52 bpm

Figure

This image show a Pulse Oximeter with 98% oxygen saturation level and pulse/HR of 52 bpm. Contributed by Tripthi M. Mathew, MD, MPH, MBA, PhD

Monitor showing mixed venous oxygen saturation value (SvO2) in blue and cardiac output (CO) value in orange

Figure

Monitor showing mixed venous oxygen saturation value (SvO2) in blue and cardiac output (CO) value in orange. Contributed by Vaibhav Bora, MD

Mixed venous oxygen saturation

Figure

Mixed venous oxygen saturation. Contributed by Chan Lee

References

1.

Kaufman DP, Kandle PF, Murray I, Dhamoon AS. StatPearls [Internet]. StatPearls Publishing; Treasure Island (FL): Aug 6, 2021. Physiology, Oxyhemoglobin Dissociation Curve. [PubMed: 29762993]

2.

Clause D, Detry B, Rodenstein D, Liistro G. Stability of oxyhemoglobin affinity in patients with obstructive sleep apnea-hypopnea syndrome without daytime hypoxemia. J Appl Physiol (1985). 2008 Dec;105(6):1809-12. [PubMed: 18948445]

3.

Hanning CD, Alexander-Williams JM. Pulse oximetry: a practical review. BMJ. 1995 Aug 05;311(7001):367-70. [PMC free article: PMC2550433] [PubMed: 7640545]

4.

Bongard F, Sue D. Pulse oximetry and capnography in intensive and transitional care units. West J Med. 1992 Jan;156(1):57-64. [PMC free article: PMC1003148] [PubMed: 1734600]

5.

Hinkelbein J, Koehler H, Genzwuerker HV, Fiedler F. Artificial acrylic finger nails may alter pulse oximetry measurement. Resuscitation. 2007 Jul;74(1):75-82. [PubMed: 17353079]

6.

Grace RF. Pulse oximetry. Gold standard or false sense of security? Med J Aust. 1994 May 16;160(10):638-44. [PubMed: 8177111]

7.

Pu LJ, Shen Y, Lu L, Zhang RY, Zhang Q, Shen WF. Increased blood glycohemoglobin A1c levels lead to overestimation of arterial oxygen saturation by pulse oximetry in patients with type 2 diabetes. Cardiovasc Diabetol. 2012 Sep 17;11:110. [PMC free article: PMC3489581] [PubMed: 22985301]

8.

Sarikonda KV, Ribeiro RS, Herrick JL, Hoyer JD. Hemoglobin lansing: a novel hemoglobin variant causing falsely decreased oxygen saturation by pulse oximetry. Am J Hematol. 2009 Aug;84(8):541. [PubMed: 19536848]

9.

Lee WW, Mayberry K, Crapo R, Jensen RL. The accuracy of pulse oximetry in the emergency department. Am J Emerg Med. 2000 Jul;18(4):427-31. [PubMed: 10919532]

10.

American Thoracic Society; American College of Chest Physicians. ATS/ACCP Statement on cardiopulmonary exercise testing. Am J Respir Crit Care Med. 2003 Jan 15;167(2):211-77. [PubMed: 12524257]